Abstract:
A new selective and sensitive high-performance liquid chromatography coupled with tandem mass spectrometry (LC-MS/MS) method was developed for the quantification of natamycin in rabbit tears using amphotericin B as internal standard (IS). Chromatographic separation was achieved on a Luna Cyano column (100 mm × 2 mm, 3µm) using ammonium acetate buffer (pH 4; 3.5 mM) : methanol (10:90, v/v) as the mobile phase. The run time was 5 min. Detection was performed by negative ion electrospray ionization in multiple reaction monitoring (MRM) mode. The calibration curve was linear over the concentration range from 25 to 800 ng/ml, and lower limit of detection of 12.5 ng/ml. The accuracy and precision of the method were within the acceptable limit of ± 20% at the lower limit of quantitation and ± 15% at other concentrations. Natamycin was stable during the battery of stability studies viz., bench-top, auto-sampler, freeze/thaw cycles and 30 days storage in a freezer at -70 ± 10 °C. The method was successfully applied to the ocular pharmacokinetic studies of natamycin eye drops in New Zealand rabbit tears